Harga bagus  on line

detail produk

Rumah > Produk >
Sumber Sinar-X Hamamatsu
>
Hamamatsu X Ray Source Providing Detailed X Ray Imaging for Control of Metal Components Printed Circuit Boards

Hamamatsu X Ray Source Providing Detailed X Ray Imaging for Control of Metal Components Printed Circuit Boards

Nama Merek: HAMAMATSU
Nomor Model: L11831-01
Moq: 1
Ketentuan Pembayaran: T/T
Informasi Detail
Tempat asal:
JEPANG
Nama merek:
Hamamatsu
Tempat asal:
JEPANG
Kemasan rincian:
Karton
Bahan Sasaran:
Tungsten
Komunikasi:
Antarmuka: RS-232C (konektor D-sub 9-pin)
Menyoroti:

Hamamatsu X-ray source for metal inspection

,

X-ray imaging for printed circuit boards

,

detailed X-ray source with warranty

Deskripsi produk
Hamamatsu X-Ray Source: Detailed X-Ray Imaging for Metal Components & Printed Circuit Boards
90 kV MICROFOCUS X-RAY SOURCE
Applications
  • Non-destructive Inspection
  • X-ray CT
  • In-line X-ray Inspection
Applicable Objects
  • Electronic components
  • Printed circuit boards
  • Plastic components
  • Metal components
Key Features
  • Focal Spot Size: 15 μm (at 90kV)
  • High Precision: 5 μm focal spot enables sharp, clear X-ray images even at high magnification
  • Easy Handling: Fully operable from an external PC
  • Serial Port Control: RS-232C interface
  • Integrated Design: One package includes sealed X-ray tube, high-voltage power supply, and control function
Tube Current Operational Range
Hamamatsu X Ray Source Providing Detailed X Ray Imaging for Control of Metal Components Printed Circuit Boards 0
Dimensional Outline
Hamamatsu X Ray Source Providing Detailed X Ray Imaging for Control of Metal Components Printed Circuit Boards 1
Technical Specifications
GENERAL SPECIFICATIONS
Input Voltage (DC) +24 V
Power Consumption (Max.) 96 W
Operating Ambient Temperature +10 to +40
Storage Temperature 0 to +50
Operating and Storage Humidity Below 85% (No Condensation) %
Weight Approx. 10 kg
Conformance Standards CE (EMC: IEC61326-1, Group 1, Class A)
Operation Continuous
High Voltage Power Supply Built-in
X-RAY TUBE SPECIFICATIONS
X-ray Tube Sealed Type
X-ray Tube Cooling Method Convection Cooling
X-ray Tube Window Material/Thickness Beryllium/150 μm
Target Material Tungsten
Tube Voltage Operational Range 20 to 90 kV
Tube Current Operational Range 10 to 200 (8 W Max.) μA
Maximum Output 8 W
X-ray Focal Spot Size 7 (5μm at 4W) μm
X-ray Beam Angle (Coned) 39 degrees
Focus to Object Distance (FOD) 9.5 mm
X-RAY CONTROL PART
Function Tube Voltage and Tube Current Preset / Auto Warm-up
Protection Interlock
External Control RS-232C
Applicable OS Windows 2000 Professional, XP Professional
Computer Operating Conditions CPU: Intel Pentium or Higher, Memory: 64MB or More