Product Description
Hamamatsu X Ray Source 130 kV Microfocus X Ray Source for Metal Components with Focal Spot Size of 5 μm
Product Description

X-RAY SOURCE 130 kV MICROFOCUS X-RAY SOURCE L9181-02

Key Features
  • Focal spot size: 5μm (at 4W) - Offers sharp and clear X-ray images even at high magnification
  • No high voltage cable connection required - High voltage power supply is integrated
  • External control via RS-232C interface
  • High power: Maximum output 39 W
Applications
Non-destructive inspection
X-ray CT

Applicable objects:

  • Electronic component
  • Printed circuit board
  • Plastic component
  • Metal component
Specifications
GENERAL
Parameter Description/Value Unit
X-ray tube voltage setting range 0 to 130 kV
X-ray tube current setting range 0 to 300 μA
X-ray tube voltage operational range 40 to 130 kV
X-ray tube current operational range 10 to 300 μA
Maximum output Small focus mode: 8 W
Middle focus mode: 16 W
Large focus mode: 39 W
W
X-ray focal spot size (Nominal value) Small focus mode: 8 (5μm at 4 W)
Middle focus mode: 20
Large focus mode: 40
μm
X-ray output window material/Thickness Beryllium/0.2 mm
X-ray beam angle Approx.45 degree
Focus to object distance (FOD) Approx.13 mm
Target material Tungsten
Weight Approx.10.5 kg
Communication method Interface: RS-232C (9-pin D-sub connector)
RATINGS
Parameter Description/Value Unit
Input voltage (DC) +24 (+2.4,-0) V
Power consumption Less than 120 W
Read output Continuous rating --
Operating ambient temperature +10 to +40
Storage ambient temperature 0 to +50
Operating and storage humidity 20 to 85 (No condensation) %
REGULATION AND STANDARD
Parameter Description Unit
RoHS Directive EN 50581 Category 9
EMC IEC/EN 61326-1
Emission limits: CISPR 11 Group 1 Class A
Immunity requirements: Table 2
CONTROL SOFTWARE
Parameter Description Unit
Applicable PC PC/AT compatible
Applicable OS Windows® XP, 7
Interface RS-232C

NOTE:

① See the graph of the X-ray tube voltage/current operation range.

② Reference value: With 50% of maximum X-ray emission.

③ This weight includes the accessories of approx. 0.25 kg.

④ The control software is provided as a sample software for the purpose of MFX operation. The performance of the software is not guaranteed.